• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/03/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that post-operatively to a successful cryo ablation procedure, it was noted that blood leaked from the valve of the sheath.The case was completed with cryo.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the sheath was returned and analyzed.Visual inspection of the sheath 4fc12 lot number 46062 showed the device was intact with no apparent issues.Flushing and air aspiration testing did not show any air passing through the tube or expelled from the distal tip.Functional testing of the sheath confirmed the hemostatic valve was leaking, and air bubbles were observed through the valve.It was suspected the valve disk is torn.In conclusion, the reported issue of hemostatic valve leak during aspiration was confirmed through product testing.The reported sheath 4fc12 lot number 46062 failed the returned product inspection due to a leaking hemostatic valve.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
lisa robertson
8200 coral sea st ne
mounds view, MN 55112
7635262723
MDR Report Key8202701
MDR Text Key131712666
Report Number3002648230-2018-00970
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number46062
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/26/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/02/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-