Additional information : device manufactured between: 05/15/2018 to 05/17/2018.Four devices were received for evaluation.A visual inspection with the unaided and with magnification revealed a tear at the lower right corner edge on sample 1.Upon functional testing with tap water a leak was observed.The reported condition was verified.The cause of condition could not be determined.Functional testing of samples 2, 3, and 4 revealed leaks at the spike port caps and no leaks were observed in the corner of the bags.Functional testing was performed by filling the bags with tap water.The cause of the leaks could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.This issue is being further investigated.Should additional relevant information become available, a supplemental report will be submitted.
|