• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE ¿ ROUND LAKE 2000 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE ¿ ROUND LAKE 2000 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938740
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 11/10/2018
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that four (4) eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bags were leaking at the corner.The leaks were discovered during setup/preparation.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information : device manufactured between: 05/15/2018 to 05/17/2018.Four devices were received for evaluation.A visual inspection with the unaided and with magnification revealed a tear at the lower right corner edge on sample 1.Upon functional testing with tap water a leak was observed.The reported condition was verified.The cause of condition could not be determined.Functional testing of samples 2, 3, and 4 revealed leaks at the spike port caps and no leaks were observed in the corner of the bags.Functional testing was performed by filling the bags with tap water.The cause of the leaks could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.This issue is being further investigated.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2000 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE ¿ ROUND LAKE
round lake IL
Manufacturer (Section G)
BAXTER HEALTHCARE ¿ ROUND LAKE
25212 w. illinois route 120
round lake IL 60073
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8203964
MDR Text Key131701141
Report Number1416980-2018-08068
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477244
UDI-Public(01)00085412477244
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date04/30/2021
Device Catalogue NumberH938740
Device Lot Number60126015
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2019
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/03/2018
Initial Date FDA Received12/28/2018
Supplement Dates Manufacturer Received01/30/2019
Supplement Dates FDA Received02/20/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-