Model Number CD3367-40QC
Device Problem Over-Sensing (1438)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/18/2018
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented in clinic for a routine follow up.It was noted that the implantable cardioverter defibrillator exhibited multiple non sustained ventricular oversensing due to t wave oversensing.The cause was thought to result from a turned off low frequency attenuation (lfa) filter.Recommendations were made to turn the lfa on at the next appointment.The patient was stable.
 
Event Description
New information received notes this was post paced t wave oversensing.
 
Event Description
New information states that the device was reprogrammed.The patient stable with no further episodes of oversensing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8204008
MDR Text Key131709623
Report Number2017865-2018-19736
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 03/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2019
Device Model NumberCD3367-40QC
Device Lot NumberP000042100
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/13/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/19/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-