MAUDE Adverse Event Report: BAXTER HEALTHCARE ¿ ROUND LAKE 3000 ML TPN BAG; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE ¿ ROUND LAKE 3000 ML TPN BAG; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938741
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Device manufacture date: the lot was manufactured between 12/02/2017 - 12/14/2017.The device was received for evaluation.The bottom right corner was cut where the customer likely drained the contents.Visual and magnified inspections were performed and revealed a tear/hole in the top seam of the returned sample.Functional testing was performed and leak was observed.The cause of the condition could not be determined.This issue is being further investigated.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
During sample evaluation, a 3000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag leaked from the weld.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3000 ML TPN BAG
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE ¿ ROUND LAKE
round lake IL
Manufacturer (Section G)
BAXTER HEALTHCARE ¿ ROUND LAKE
25212 w. illinois route 120
round lake IL 60073
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8204994
MDR Text Key131701089
Report Number1416980-2018-08098
Device Sequence Number1
Product Code LHI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 12/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberH938741
Device Lot Number60101131
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2018
Initial Date FDA Received12/28/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-