SMITH & NEPHEW, INC. JRNY UNI TIBIAL IMPACTOR; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 71441353
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/30/2018
Event Type  malfunction  
Event Description
It was reported that during tka the tip broke off upon case.No delay, injury or back up reported.
 
Manufacturer Narrative
The associated journey uni tibial impactor was returned and evaluated.A visual inspection of the returned device confirmed the failure mode as the plastic tip on the end of the impactor bumper has fractured off, the broken piece is not returned.The device was manufactured in 2008 and it exhibits signs of significant wear/usage.The crack is likely due to repeated impacts seen during impaction.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY UNI TIBIAL IMPACTOR
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8205528
MDR Text Key131706633
Report Number1020279-2018-02949
Device Sequence Number1
Product Code HSX
UDI-Device Identifier03596010588050
UDI-Public03596010588050
Combination Product (y/n)N
PMA/PMN Number
K113038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71441353
Device Lot Number08DM05453
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/24/2018
Date Manufacturer Received11/30/2018
Is This a Reprocessed and Reused Single-Use Device? No