SMITH & NEPHEW, INC. JRNY UNI TIBIAL IMPACTOR; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Catalog Number 71441353 |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/30/2018 |
Event Type
malfunction
|
Event Description
|
It was reported that during tka the tip broke off upon case.No delay, injury or back up reported.
|
|
Manufacturer Narrative
|
The associated journey uni tibial impactor was returned and evaluated.A visual inspection of the returned device confirmed the failure mode as the plastic tip on the end of the impactor bumper has fractured off, the broken piece is not returned.The device was manufactured in 2008 and it exhibits signs of significant wear/usage.The crack is likely due to repeated impacts seen during impaction.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|
|