• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PROMUS PREMIER; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PROMUS PREMIER; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number 24623
Device Problems Break (1069); Difficult to Remove (1528)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/01/2018
Event Type  malfunction  
Manufacturer Narrative
Device is a combination product.
 
Event Description
It was reported that catheter removal difficulties were encountered and shaft break occurred.The target lesion was located in a 3mm artery in the leg.After a 0.014 non-bsc guide wire was advanced, a 3.50 x 28 promus premier drug-eluting stent was advanced to treat the lesion.However, after 20cm of progression, the device can no longer advance.It was impossible to move the device backward or forward resulting to the ablation of the guide and rupture of the shaft.The procedure was completed with dilation without stenting.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS PREMIER
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key8205954
MDR Text Key131702347
Report Number2134265-2018-64709
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 12/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/19/2019
Device Model Number24623
Device Catalogue Number24623
Device Lot Number0020814669
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/12/2018
Initial Date FDA Received12/31/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/19/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
XT 0.014 BIOTRONIK GUIDE WIRE
-
-