SMITH & NEPHEW, INC. LEGION TRL STEM CONNECTION ROD; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLY
|
Back to Search Results |
|
Catalog Number 71434317 |
Device Problem
Fitting Problem (2183)
|
Patient Problem
Injury (2348)
|
Event Date 12/04/2018 |
Event Type
malfunction
|
Event Description
|
It was reported that during surgery this instrument will not fit in to the t-handle properly.No delay was reported.No backup device was available.The procedure finished with vice grips.
|
|
Manufacturer Narrative
|
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The chuck end of the device was flared out from impact damage and would not engage a mating part as intended.The device was manufactured in 2013 and exhibits signs of significant wear/usage.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
|
|
Search Alerts/Recalls
|
|
|