• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problems Mechanical Problem (1384); Decrease in Pressure (1490)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/12/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluation by manufacturer: a field service engineer (fse) contacted the customer over the phone to address the reported event.During troubleshooting, the customer informed fse that they had set the flow factor to 1.57 ml/min, which increased the pump pressure.This is what caused the leak.The customer adjusted the flow factor back to 1.12 ml/min and the pressure went back to normal.The instrument was working as expected.No further action was required by the fse.A 13-month complaint history review and service history review on g8 serial number (b)(4) from 12-nov-2017 through (b)(6) 2018.There were three similar reported issues, including this event, found during the searched period.The g8 operator's manual under chapter 3, assay operations, section 7, states the following: other items to check: check the flow line connections, particularly the filter and the column inlet and outlet for leaks during warming up operations.Tighten the connection if a leak is found.The g8 variant analysis mode training manual under lesson 8 - troubleshooting, states the following: adjusting the flow rate - how and why: on the tosoh automated glycohemoglobin analyzer hlc-723g8; variant analysis mode it may be necessary to adjust the flow rate because either unidentifiable peaks appear on all the chromatograms or the average retention time for various peaks has changed significantly.The flow rate is changed by changing the flow factor in the instrument.The flow factor is generally 1.00 ml/min.The flow factor should only be adjusted +/- 0.05 of the default factory setting.The most probable cause of the reported issue was due to incorrect flow factor setting by the customer which lead to increased pump pressure and the reported leak.
 
Event Description
A customer reported that the prefilter assembly at the peak tubing was leaking on the g8 instrument.The customer stated that the wing nut was broken which resulted in low pressure.The system was down.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay of results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key8209426
MDR Text Key132537375
Report Number8031673-2018-05382
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131580
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 01/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/02/2019
Distributor Facility Aware Date12/12/2018
Device Age6 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer01/02/2019
Initial Date Manufacturer Received 12/12/2018
Initial Date FDA Received01/02/2019
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-