• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. HIP KIT; LAVAGE, JET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. HIP KIT; LAVAGE, JET Back to Search Results
Catalog Number 00515048201
Device Problem Expulsion (2933)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/20/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Customer has indicated that product is in the process of being returned to zimmer biomet for investigation.Once the investigation is complete, a follow-up mdr will be submitted.
 
Event Description
It was reported that during the surgery, the battery has already exploded when it was opened.Alternate device was used to complete the procedure.There was a minimal delay of (0-15 minute) to prepare an alternative product.No adverse events were reported as a result of this malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: event problem: battery explosion.Device history record (dhr) for z000011159 lot number 00515048201, review noted no related non-conformances, requests for deviation (rfd), change notices (cn), or any other issues with manufacturing.The dhr review found no issues with the device and all verifications, inspections, and tests were successfully completed.On (b)(6) 2018, it was reported from (b)(6) hospital that the battery has already exploded since it was opened.On 18 february 2019, a returned product investigation was performed on the 00515048201.The physical evaluation revealed that the battery had ruptured inside the packaging.The results of the returned product investigation have confirmed the reported event.While the returned product investigation confirmed that the 00515048201 battery had ruptured inside of the packaging, it cannot be determined from the information provided what actually caused the reported event.Therefore, based on the information provided, a specific root cause of the reported event cannot be determined.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.
 
Event Description
No additional event information was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIP KIT
Type of Device
LAVAGE, JET
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
MDR Report Key8210368
MDR Text Key131895503
Report Number0001526350-2019-00001
Device Sequence Number1
Product Code FQH
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/02/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2021
Device Catalogue Number00515048201
Device Lot NumberZ000011159
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2019
Was the Report Sent to FDA? No
Date Manufacturer Received03/13/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-