• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC ELECTRIC PEN DRIVE 60,000 RPM; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC ELECTRIC PEN DRIVE 60,000 RPM; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 05.001.010
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/16/2018
Event Type  malfunction  
Manufacturer Narrative
The manufacturing location is currently not available.Service review: a review of the service history record indicates that the device has been serviced within the last year for a service condition that is not relevant to the current reported condition.Device evaluation: the actual device was returned for evaluation.The device was evaluated and the reported condition was confirmed.A visual and functional assessment was performed on the device which found that the device started to run and then it stopped when it was getting warm.It was determined that the device failed for check function and direction of rotation, check function with test hand switch and check function with foot pedal.During service/repair, it was observed that the internal electronic components were damaged.The assignable root cause was determined to be due to normal wear out from use.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Event Description
It was reported that during pre-surgery or set up, it was observed that the electric pen drive device had no power and was not working.During in-house engineering evaluation, it was observed that the device started to run and then it stopped when it was getting warm.The event was not related to surgery.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTRIC PEN DRIVE 60,000 RPM
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
N/I
n/i
n/i
Manufacturer Contact
kara ditty-bovard
eimattstrasse 3
oberdorf 4436
SZ   4436
6103142063
MDR Report Key8214166
MDR Text Key132211907
Report Number8030965-2018-59346
Device Sequence Number1
Product Code GEY
UDI-Device Identifier7611819167711
UDI-Public(01)7611819167711(11)100324
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 10/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2018
Initial Date Manufacturer Received 12/17/2018
Initial Date FDA Received01/03/2019
Date Device Manufactured03/24/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-