Model Number 26926
Device Problems Entrapment of Device (1212); Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/12/2018
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: older than 18 years.
 
Event Description
It was reported that the stent became stuck on the guidewire.An innova stent was selected for use for a lower limb stent procedure in the left superficial femoral artery (sfa).The target lesion was moderately tortuous, severely calcified and 100% stenosed.During procedure, the stent delivery system became stuck to a non-bsc guidewire and the delivery shaft became stretched.The stent was then removed with the guidewire together, and the procedure was completed with another innova stent.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOVA VASCULAR
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
two scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key8214561
MDR Text Key132046708
Report Number2134265-2018-64608
Device Sequence Number1
Product Code NIP
UDI-Device Identifier08714729873686
UDI-Public08714729873686
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P140028
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 01/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2018
Device Model Number26926
Device Catalogue Number26926
Device Lot Number0018733069
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/17/2018
Initial Date FDA Received01/03/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/16/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-