The device history record review confirms that the device met all material, assembly and performance specifications.Visual and microscopic inspection of the device found that the stent delivery wire (sdw) was kinked at 10cm & 33 cm from the distal end, likely due to handling.The stent was returned in its deployed state and there was some slight deformation noted to the stent.No other anomalies were noted.Functional test could not be performed as the stent was returned deployed.It is possible that stent prematurely deployed outside the patient most likely due to inability to pull back the stent after unsuccessful advancement during the procedure.It is likely that procedural factors contributed to the reported and observed damages limiting the performance of the stent during the clinical procedure.Therefore, an assignable cause of component failure has been assigned to this investigation.
|