Model Number CBP241 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
Blood Loss (2597)
|
Event Date 12/06/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
Inspection of the returned pixie reservoir indicated the pressure relief valve cap was still installed, potentially contributing to poor drainage/lines popping off.Additional visual inspection showed no outward signs of any damage including the vent port or any assembly errors.Investigation is ongoing.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that during use, the customer reported that when the patient went on bypass, the affinity pixie oxygenator/cvr drainage was very poor.One of the vent lines popped off the reservoir twice and caused blood loss.The volume of blood loss (ml) was not provided.The product was replaced.They came off bypass for about 3 minutes to change out reservoir.Cpr was needed at this time.The customer stated the only known patient impact was blood loss, patient status or other details were not provided.The replacement reservoir was used to complete the case.
|
|
Manufacturer Narrative
|
Investigation conclusion: complaint is not confirmed for the pixie oxygenator/cvr's poor drainage.Review of this unit's device history record found no abnormalities or ncmrs initiated during manufacturing that would cause or contribute to the reported event.The returned device pressure relief valve cap was still installed.The product ifu instructs the user to remove the obturator cap from the cvr positive/negative pressure relief valve and to ensure the cvr is vented to the atmosphere by removing the vent port cap during setup.Using the cvr without removing the obturator cap can contribute to poor drainage and vent lines popping off.Root cause is undetermined.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|