BAXTER HEALTHCARE - MALTA CLEARLINK LUER ACTIVATED VALVE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number EMC9182
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/11/2018
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a non-baxter infusion pump presented a continuous downstream occlusion alarm during use with a clearlink iv access valve.The valve was replaced with an unspecified cap to resolve the issue.This occurred during set up.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information : the actual device was not available; therefore, a device analysis could not be completed on the actual sample.However, four (4) companion samples were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A blockage test was performed on the samples and no blockages were noted.A pump functionality test was performed and no issues were noted.The reported condition was not verified on the companion samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK LUER ACTIVATED VALVE
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE - MALTA
marsa
Manufacturer (Section G)
BAXTER HEALTHCARE - MALTA
a47 industrial estate
marsa
MT  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8217869
MDR Text Key132097443
Report Number1416980-2018-08187
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberEMC9182
Device Lot Number17H11V548
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/12/2018
Initial Date FDA Received01/04/2019
Supplement Dates Manufacturer Received01/09/2019
Supplement Dates FDA Received02/01/2019
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/27/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
ARGUS 717V INFUSION PUMP
-
-