• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONVATEC DOMINICAN REPUBLIC INC ACTIVELIFE; BAG, URINARY, ILEOSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONVATEC DOMINICAN REPUBLIC INC ACTIVELIFE; BAG, URINARY, ILEOSTOMY Back to Search Results
Model Number 650832
Device Problem Malposition of Device (2616)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number: (b)(4).
 
Event Description
It was reported the starter hole was off center in five wafers.No harm reported.Photos depicting the reported complaint issue were provided by the complainant.
 
Event Description
To date no additional patient or event details has been received.
 
Manufacturer Narrative
Lot number 8c04926 was manufactured on 04/10/2018 in the convex 1pc line.A batch record review was performed on (b)(6)2019 to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bill of materials (bom) and all the tooling information documented was also correct, under international commodity code (icc) 650832 material identification number 1211902 and manufacturing order 1401230.The batch record review supports that there were no discrepancies related to the issue reported.Photograph associated with this case is available for evaluation.Based on the investigation findings, the root cause identified for the issue ¿wafer off center in the pouch¿, is attributed to: the investigative process concluded that all the machinery/ tooling items used in the manufacturing process complied when compared against drawing and process instruction specifications; however, as part of this investigation on may 10, 2019, maintenance and facilities manager and senior process engineer performed a deep assessment to the sub-assembly station of the convex 1pc machine identifying the conditions causing the wafer off center and the possible solutions to correct the failure.As conclusion of the assessment, the chain that coordinates flange loading station with adhesive loading station needed lubrication and the chain and other components of the mechanism needed a corrective maintenance.In order to maintain this condition in the long term, a deeper intervention of the mechanism is required.Furthermore, regarding the rotary table station of the convex 1pc machine, it was identified that due to the demands of the process, the tooling (cup pouch uro 177a) used to manufactured the urodress product family require a redesign modification to reduce the variability within the manufacturing process.As a contributor factors the following opportunities for improvements were identified: reference in the process instructions ¿quality instructions¿ the applicable qc tooling used for product¿s quality inspection purpose of pouch uro 177a products family in order to address manufacturing personnel the use of them.Identification of the defect during the manufacturing process, it should be considered the implementation of a job aid for product centralization out of specification.Introduce quality control (qc) tooling used for product¿s quality inspection purpose in the calibration program to guarantee measurement assurance.There are not manpower causes associated with the root cause; even though, opportunity to improve the loading of the adhesive disc during the manufacturing process was identified by implementing a standard work instructions for convex 1pc to indicate manufacturing operators how to properly load the adhesive disc.No issues were identified for material and environment investigations.A non-conformance was created and is now complete.This issue will be monitored through the post market product monitoring review process.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 9618003.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVELIFE
Type of Device
BAG, URINARY, ILEOSTOMY
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal
MDR Report Key8218535
MDR Text Key132212883
Report Number9618003-2019-00164
Device Sequence Number1
Product Code EXH
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date04/09/2023
Device Model Number650832
Device Lot Number8C04926
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 12/18/2018
Initial Date FDA Received01/04/2019
Supplement Dates Manufacturer Received07/22/2019
Supplement Dates FDA Received08/06/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-