MAUDE Adverse Event Report: ENDOLOGIX AFX; SUPRARENAL AORTIC EXTENSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX AFX; SUPRARENAL AORTIC EXTENSION Back to Search Results
Model Number A34-34/C80-O20
Device Problems Leak/Splash (1354); Detachment of Device or Device Component (2907); Material Integrity Problem (2978); Unintended Movement (3026)
Patient Problem Failure of Implant (1924)
Event Date 09/15/2018
Event Type  Injury  
Manufacturer Narrative
The manufacturing lot evaluation confirmed all devices met specifications prior to release.The device was not returned as the device remains implanted therefore no evaluation was completed.Clinical had substantial evidence to support the following additional findings of implant separation, stent migration and stent cage dilation (3b endoleak) of the suprarenal stent graft extension (cuff) that was implanted on (b)(6) 2011 during the index procedure.The type 3b endoleak is mostly likely device related due to the use of strata material.A root cause investigation was carried out for all afx complaints having an identified failure mode of a type iiib endoleak.Endologix implemented the following corrective actions with the intent of reducing type iiib endoleak events; 1.Upgraded graft material (i.E.Duraply) and 2.Updates to the ifu and additional physician training.The change to duraply graft material and the ifu changes were put in place july 2014.No additional investigation of this reported event is planned.Endologix will continue to monitor this complaint and similar complaints in the event further investigation is needed.
 
Event Description
At the completion of the clinical assessment on 12/06/2018, for a previous report event (2031527-2018-00886), clinical was able to find substantial evidence to support the following additional findings of implant separation, stent cage dilation and stent migration of the suprarenal stent graft extension implanted on (b)(6) 2011.This complaint is for the suprarenal stent graft extension implanted on (b)(6) 2011 only,.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Type of Device
SUPRARENAL AORTIC EXTENSION
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer Contact
victor arellano
2 musick
irvine, CA 92618
9495984671
MDR Report Key8220432
MDR Text Key132194379
Report Number2031527-2019-00012
Device Sequence Number1
Product Code MIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P040002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2012
Device Model NumberA34-34/C80-O20
Device Lot NumberW11-3592-030
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/06/2018
Initial Date FDA Received01/04/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/10/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age85 YR
-
-