STRYKER INSTRUMENTS-KALAMAZOO ZYPHR DISPOSABLE CRANIAL PERFORATOR, LARGE 14/11MM; DRILLS, BURRS, TREPHINES & ACCESSORIES Back to Search Results
Catalog Number 5100060001
Device Problem Failure to Auto Stop (2938)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/13/2018
Event Type  malfunction  
Manufacturer Narrative
Awaiting device evaluation.A follow up report will be filed once the quality investigation is complete.
 
Event Description
It was reported that during a cranial procedure, the perforator bit did not disengage.It was also reported that there was no adverse consequences as a result of this event.It was further reported that there was no surgical delay and the procedure was completed successfully.
 
Manufacturer Narrative
The perforator product reported involved with this event was returned for evaluation and the disengagement mechanism within the perforator was tested and functioned as intended.The reported failure of failure to disengage could not be confirmed.The ifu #5100-060-700 revision f provides instructions on how to test the disengagement function of the perforator prior to use.The quality investigation is complete.
 
Event Description
It was reported that during a cranial procedure, the perforator bit did not disengage.It was also reported that there was no adverse consequences as a result of this event.It was further reported that there was no surgical delay and the procedure was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZYPHR DISPOSABLE CRANIAL PERFORATOR, LARGE 14/11MM
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-IRELAND
instruments division
carrigtwohill bus. & tech park
carrigtwohill NA
Manufacturer Contact
una barry
instruments division
carrigtwohill bus. & tech park
carrigtwohill NA
214532900
MDR Report Key8222655
MDR Text Key132294082
Report Number0001811755-2019-00034
Device Sequence Number1
Product Code HBF
UDI-Device Identifier04546540716224
UDI-Public04546540716224
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/07/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/01/2021
Device Catalogue Number5100060001
Device Lot Number18138027
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2019
Is the Reporter a Health Professional? No
Date Manufacturer Received02/08/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/18/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-