On (b)(6) 2018, a 23mm mechanical valve (standard) was implanted in the aortic position in a patient with complex congenital heart disease.The patient had undergone multiple previous operations including a konno procedure that had now calcified.After discharge, the patient was not compliant with prescribed anticoagulation therapy and at the time of readmission, the mechanical valve had evidence that it was now thrombosed and required intervention.The patient was placed on tandem heart lvad to support stable hemodynamic status and treated until (b)(6) 2018 when the patient was brought to the operating room.At surgery, tissue and debris were removed from the valve; however the valve remained implanted.Post-operatively, the patient is reported to be stable.
|
An event of thrombus was reported in a patient with known non-compliance with the prescribed anticoagulation regiment.As the device remains implanted, no device analysis is available.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident is related to the thrombus formation on the valve in a non-compliant patient.
|