(b)(4).Device evaluation by manufacturer: a field service engineer (fse) contacted the customer over-the-phone to address the reported event.During troubleshooting, the customer was unable to resolve the issue.However, the customer called back stating that they cleaned the contacts for the offline/online lever of the thermal printer.This resolved the issue.The printer was operating as expected.There was no further action required by fse.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from 10-nov-2017 through aware date (b)(4) 2018.There were no similar complaints found during the searched period.The g8 variant analysis mode operator's manual under chapter 6, troubleshooting states the following: 6.3 error messages when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.401 printer off line: the printer paper holding lever is lifted up.Set the lever correctly.The most probable cause of the reported event was due to a dirty contacts on the offline/online lever of the thermal printer.
|
A customer reported getting a 401 printer off line error message on the g8 instrument.The customer tried reinstalling paper, moved release lever to the correct position and powered the instrument off and on.The problem persisted.The instrument was down.A field service engineer (fse) was dispatched to address the reported event, which could have resulted in delay of results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|