Model Number G8
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/10/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluation by manufacturer: a field service engineer (fse) contacted the customer over-the-phone to address the reported event.During troubleshooting, the customer was unable to resolve the issue.However, the customer called back stating that they cleaned the contacts for the offline/online lever of the thermal printer.This resolved the issue.The printer was operating as expected.There was no further action required by fse.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from 10-nov-2017 through aware date (b)(4) 2018.There were no similar complaints found during the searched period.The g8 variant analysis mode operator's manual under chapter 6, troubleshooting states the following: 6.3 error messages when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.401 printer off line: the printer paper holding lever is lifted up.Set the lever correctly.The most probable cause of the reported event was due to a dirty contacts on the offline/online lever of the thermal printer.
 
Event Description
A customer reported getting a 401 printer off line error message on the g8 instrument.The customer tried reinstalling paper, moved release lever to the correct position and powered the instrument off and on.The problem persisted.The instrument was down.A field service engineer (fse) was dispatched to address the reported event, which could have resulted in delay of results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key8225029
MDR Text Key133934031
Report Number8031673-2018-05370
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131580
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 01/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/07/2019
Distributor Facility Aware Date12/10/2018
Device Age8 YR
Event Location Hospital
Date Report to Manufacturer01/07/2019
Initial Date Manufacturer Received 12/10/2018
Initial Date FDA Received01/07/2019
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-