(b)(4).Batch # p93a23.Investigation summary: the device was returned with the blade tip broken off and returned.During functional testing on gen11 an alert screen was displayed.A probable cause for the device to stop activating and the gen11 to display an alert screen is blade damage.The device will stop activating when the blade becomes damaged.The device was disassembled to inspect the blade and the blade breaking point was located at an area inside the tube proximal to the tissue pad.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result in failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result are such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damage blade can result in a broken blade.No conclusion could be reached as to what cause the damage blade.The damaged blade found is not related to the tissue effect issues reported.Due to the returned condition of the device we are unable to investigate further the reported issue.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
|