(b)(4).Device evaluation by manufacturer: a field service engineer (fse) visited the customer to address the reported event.During servicing, thoroughly cleaned the large and small drive screws and guide rods with alcohol and then lubricated them.Fse then ran quality controls and patient samples with acceptable results.However, after performing a periodic maintenance (pm), the instrument got a 710 error message.Fse verified the problem and cleaned the z guide rod and drive screw again and lubricated them.Fse also replaced the sample needle.Fse found that the column had been in use with over 2500 injections.Fse then reran quality controls and patient samples with acceptable results.The instrument was operating as expected.There was no further action required by fse.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from 19-nov-2017 through aware date (b)(4) 2018.There were no similar complaints identified during the searched period.The g8 variant analysis mode operator's manual under chapter 1 states: 1.4 storage and stabiilty: columns are warranted for 2500 injections.Chapter 6, troubleshooting, states the following: 706 syringe-l error: explanation: operation error in syringe-l.Countermeasure: inspect x1-axis.Inspect syringe-l.Execute smp.Reset.710 z1-axis error: an abnormality occurred in the up and down movement of the sampling needle.If this occurs during a stat assay, check that the container setting (cup or tube) is correctly set.The error also occurs when the sample vial was not recognized as a primary tube, due to the disoriented sample sensor.Chapter 5, maintenance procedures, under section 5.10 provides step-by-step instructions on the sampling needle assembly replacement.The most probable cause of the reported event was the syringe drive screws were dirty, a bad sample needle, and the column injection count over 2500.
|