Model Number G8
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/18/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluation by manufacturer: a field service engineer (fse) visited the customer to address the reported event.During evaluation, the fse found that the x1-flag was out of alignment.Fse adjusted the x1 flag and ran the test loader operation without any issue.Quality controls (qc) and patient samples were run with acceptable results.The instrument was operating as expected.There was no further action required by fse.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from 18-nov-2017 through aware date (b)(4) 2018.There were no similar complaints including this complaint identified during the searched period.The g8 variant analysis mode operator's manual under chapter 6, troubleshooting states the following: 6.3 error messages: when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.708 x1-axis error: an abnormality occurred in the operation error in the x1-axis.If this occurs, inspect the x1-axis.Then, execute sl rotate.The most probable cause of the reported event was the x1 axis was out of alignment.
 
Event Description
A customer reported getting a 708 x1-axis error message on the g8 instrument.Technical support (ts) instructed the customer to clean the apertures/footers and blow canned air along the belt and rotate the belt.The error persisted.The instrument was down.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key8229669
MDR Text Key133934762
Report Number8031673-2018-05412
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131580
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 01/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/08/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/08/2019
Distributor Facility Aware Date12/18/2018
Device Age8 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer01/08/2019
Date Manufacturer Received12/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse