(b)(4).Device evaluation by manufacturer: a field service engineer (fse) visited the customer to address the reported event.During evaluation, the fse found that the x1-flag was out of alignment.Fse adjusted the x1 flag and ran the test loader operation without any issue.Quality controls (qc) and patient samples were run with acceptable results.The instrument was operating as expected.There was no further action required by fse.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from 18-nov-2017 through aware date (b)(4) 2018.There were no similar complaints including this complaint identified during the searched period.The g8 variant analysis mode operator's manual under chapter 6, troubleshooting states the following: 6.3 error messages: when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.708 x1-axis error: an abnormality occurred in the operation error in the x1-axis.If this occurs, inspect the x1-axis.Then, execute sl rotate.The most probable cause of the reported event was the x1 axis was out of alignment.
|