Investigation summary: the event unit was returned to applied medical for evaluation, along with several clips.Testing was performed on the event unit, which confirmed the complainant¿s experience of clips not closing properly.The feeder, a metal component located in the shaft, was measured and was found to be out of specification.Based on the condition of the returned unit and clips, it is likely that the reported event was caused by the feeder that was found to be out of specification upon receipt.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.Applied medical continuously seeks to improve the form, function and ease of use of its products.As part of this process, applied medical has recently implemented process and inspection enhancements intended to further minimize the potential for this type of event to occur.The event unit was manufactured prior to implementation of the process and inspection enhancements.
|