It was reported that the patient was admitted to the hospital on (b)(6) 2018 due to leg pain and syncope.Pauses were noted on telemetry in the hospital.Device interrogation revealed, the right ventricular lead exhibited low amplitude r wave signals and intermittent oversensing of noise resulting in pacing inhibition.Noise could be reproduced with pocket manipulation on both the right ventricular and left ventricular leads, the noise caused the device to both inhibit pacing and register some of this noise as tachy events.It was noted that the left ventricular lead had a history of elevated thresholds and pacing impedances and had been stable and functional with these measurements for some time.The device was reprogrammed.The patient was in stable condition during the event.The leads were explanted and replaced on (b)(6) 2019.The patient was in stable condition during and post procedure.
|