Upon visual inspection, the polyaxial screw is confirmed to be fractured along the axial plane.The fracture is superior to the center of the screw, approximately 10mm from the screw head.Beach marks can be observed on the fracture face, which is indicative of fatigue fracture.It is possible that this screw fractured due to fatigue, as it has beach marks on fracture surface and was implanted for approximately 13 to 24 months.These internal fixation devices serve as load bearing mechanisms which maintain alignment until healing has occurred.If there is a delay in healing or healing does not occur, then it is possible that the implant could eventually break, bend or loosen due to load bearing and/or residual motion.Manufacturing records were reviewed, and no relevant manufacturing issues were discovered.Per the warnings and precautions section of the product insert, internal fixation devices are load sharing devices which maintain alignment until healing occurs.If healing is delayed or does not occur the implant could eventually break, bend or loosen.
|
It was reported that an everest polyaxial screw fractured post-operatively.This fracture was discovered on an x-ray image during the patient's follow up visit.This event resulted in a revision surgery, over one year after the index surgery, which occurred on an unspecified date.
|