ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION KIT; ANESTHESIA CONDUCTION CATHETER Back to Search Results
Catalog Number AK-05502
Device Problems Difficult to Remove (1528); Difficult to Advance (2920)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/21/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
It was reported that there is difficulty feeding and removing the catheter.
 
Manufacturer Narrative
(b)(4).A device history record review was performed on the epidural catheter and needle with no relevant findings.The ifu for this kit, e-17019-109a; rev.07, was reviewed as a part of this complaint investigation.The ifu warns the end user, "never advance catheter more than 5 cm beyond the needle tip.Advancing catheter more than 5 cm increases the likelihood of catheter-related complications." the ifu also warns the user, "never tug or quickly pull on catheter during removal from patient to reduce risk of catheter breakage.Do not apply additional tension on the catheter if catheter begins to stretch excessively.Reposition patient to open the vertebral interspaces and re-attempt removal if resistance is encountered or if catheter stretches excessively during removal." a corrective action is not required at this time as a potential root cause could not be determined based upon the information provided and without a sample.Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed on the epidural catheter and needle with no evidence to suggest a manufacturing related cause.Therefore, the potential cause of the catheter being difficult to feed and remove could not be determined based upon the information provided and without a sample.
 
Event Description
It was reported that there is difficulty feeding and removing the catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION KIT
Type of Device
ANESTHESIA CONDUCTION CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key8232462
MDR Text Key132568259
Report Number1036844-2019-00021
Device Sequence Number1
Product Code BSO
Combination Product (y/n)N
PMA/PMN Number
K140110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2020
Device Catalogue NumberAK-05502
Device Lot Number23F18H0376
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/23/2018
Initial Date FDA Received01/09/2019
Supplement Dates Manufacturer Received02/07/2019
Supplement Dates FDA Received02/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-