Model Number MODEL 100
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/16/2018
Event Type  malfunction  
Manufacturer Narrative
The reported event for the autopulse platform (sn# (b)(4)) is confirmed through functional testing and archive data review.Upon visual inspection, observed damage on the front enclosure of the autopulse platform, unrelated to the reported complaint.The front enclosure has been replaced to remedy the issue.Upon powering on, the platform displayed ua45 (drive shaft not at home position) error code during initial functional testing.The drive shaft has been rotated to home position to clear the ua45 error message.The archive data review showed multiple occurrence of ua07 (discrepancy between load1 and load2 too large) and ua45 (not at "home" position after power-on/restart) error messages on the reported event date.However, ua07 is unrelated to the reported complaint.User advisory is a clearable error message and is designed into the platform to alert the operator that autopulse has detected one of several conditions.The autopulse user guide instructs to clear (ua) 45, the operator needs to pull up the lifeband until the chest bands are fully extended.This action will move the driveshaft to its home position.The user advisory will persist until the driveshaft is returned to its home position.Ua07 is an indication that the load sensing system has detected a weight/load imbalance between the two load cells.The recommended actions to take for this type of user advisory are: ensure the patient is properly aligned (armpits on the yellow line), deploy the shoulder restraint to mitigate patient movement and press restart to clear the ua.Following service, the autopulse was subjected to a 10 minutes run-in test using the 95% patient large resuscitation test fixture (lrtf) without any fault or error.The brake gap inspection was performed and verified the brake gap was within the specification.Load cell characterization test was performed and confirmed both cell modules are functioning within the specification.Historical complaints were reviewed for service information related to the reported complaint and there was no previous history of complaint reported for autopulse platform with serial number 33696.
 
Event Description
As reported, during deployment, the autopulse platform (sn (b)(4)) displayed ua45 (not at "home" position after power-on/restart) error message.The call was for cardiac arrest patient.The user attempted to troubleshoot by setting the platform back to home position, but was unable to clear the ua45 error message.No known impact or patient consequence information was available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION INC
2000 ringwood avenue
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION INC
2000 ringwood avenue
san jose CA 95131
Manufacturer Contact
kimthoa nguyen
2000 ringwood ave
san jose, CA 95131
4085411030
MDR Report Key8232770
MDR Text Key132579583
Report Number3010617000-2018-01288
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/21/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/12/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-