The instrument exhibits no outwardly damage or failure.We investigated the mechanical functions.The clamping and cutting function are working correctly.Then we investigated the cable and the hosing / handle, especially the cable entry at the handle.Here we found no abnormities like broken edges or damaged insulation.In the next step we opened the handle to check the writing.The writing is correct, no wire was trapped between the housing shells.The manufacturing documents have been checked and found to be according to specification valid during the time of production.There are no further complaints with this lot at hand.The root causes for the problem is most probably usage related.The investigation of the insulation and the writing of the instrument showed no irregularities like damaged insulations or trapped/clamped cable or wires.Because caiman is a bipolar rf system, there is no potential to earth at the instrument, so an electrical shock caused by an insulation failure of one wire is impossible.The rf current flows only between the both electrodes.This is the big advantage over the monopolar system.Without further knowledge about the circumstances we assume, that the surgeon interpreted a hepatic feedback from the ratchet in the handle (part of the clamping mechanism) as a little electric shock.No capa neccessary.Associated med watch reports: 9610612-2019-00015.
|