The reported event was unconfirmed.A silastic foley was returned.The sample was received opened and without the original packaging.The device was received in good condition.Visual evaluation noted no obvious defects.The device was inflated with 10ml water using a leur lock syringe.The balloon remained inflated for approximately one minute.No leaks were observed.The balloon was manipulated and no water leaked.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿warning: do not use ointments or lubricants having a petrolatum base.They will damage the catheter.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.Use luer tip syringe to inflate with stated ml of sterile water.Or for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.For urological use only.Do not use if package is damaged bard, bardex, and lubricath are trademarks and/or registered trademarks of c.R.Bard, inc.Single use do not resterilize manufacturer: to deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.".
|