Model Number 924256 |
Device Problem
Unstable (1667)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/09/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a healthcare provider (hcp) regarding a patient who was implanted with a neurostimulator (ins) for movement disorders.It was reported that the inner clip of the stimloc was not holding the dbs lead.After clipping the lead it pulled back a couple of cm and had to be put back in place.O arm scan at the end confirmed that the lead was relocated to the correct depth.The clip was repositioned but was still not holding the lead as well as usual.The stimloc cal was then put in to further secure the lead.The issue was resolved.There were no further complications reported.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from the manufacturer¿s representative (rep), which was confirmed with the healthcare provider (hcp), reported there was no reason for the clip not holding.The clip was still implanted and would not be returned.No further patient symptoms or complications were anticipated.
|
|