• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WAVELIGHT GMBH ALLEGRETTO WAVE EYE-Q EXCIMER LASER; OPHTHALMIC EXCIMER LASER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WAVELIGHT GMBH ALLEGRETTO WAVE EYE-Q EXCIMER LASER; OPHTHALMIC EXCIMER LASER SYSTEM Back to Search Results
Catalog Number 8065990739
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem No Code Available (3191)
Event Date 12/14/2018
Event Type  Injury  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The device history records (dhr) for the device was reviewed.The associated device was released based on company acceptance criteria.No udi required as this device was out of production prior to the september 24, 2014 udi regulation date.(b)(4).
 
Event Description
A center director reported, a patient was treated with the left eye prescription on the right eye.The patient was informed at the time of surgery and will undergo an enhancement procedure in the future.Additional information requested.
 
Manufacturer Narrative
A review of the technical service onsite history showed no abnormalities that could have contributed to this event.The laser was successfully verified prior to the day of the treatment.The root cause is user error.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEGRETTO WAVE EYE-Q EXCIMER LASER
Type of Device
OPHTHALMIC EXCIMER LASER SYSTEM
Manufacturer (Section D)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM  91058
MDR Report Key8238209
MDR Text Key132763698
Report Number3003288808-2019-00034
Device Sequence Number1
Product Code LZS
Combination Product (y/n)N
PMA/PMN Number
P020050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 03/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number8065990739
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/14/2018
Initial Date FDA Received01/10/2019
Supplement Dates Manufacturer Received03/11/2019
Supplement Dates FDA Received03/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age65 YR
-
-