On (b)(6) 2018, the (b)(6) year old patient had two septal defects for which a 35mm cribriform was selected to occlude both.The device was inserted but was not large enough to cover both defects so it was exchanged for a 18mm amplatzer septal occluder which is intended to close a defect measuring 15-17mm.The device was deployed following a push-pull test, but embolized into the left ventricle after release.The patient was immediately transferred to the operating room and the device was successfully removed.The defects were surgically closed and the patient was discharged.The physician stated the septal tissue was very thin and delicate and that the membrane may have torn during deployment.
|