• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM VA-LCP CURVED CONDYLAR PLATE/10 HOLE/230MM/RIGHT; IMPLANT,FIXATION DEVICE, CONDYLAR PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM VA-LCP CURVED CONDYLAR PLATE/10 HOLE/230MM/RIGHT; IMPLANT,FIXATION DEVICE, CONDYLAR PLATE Back to Search Results
Model Number 02.124.410
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  Injury  
Manufacturer Narrative
Additional: hrs and hwc.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on an unknown date, the patient underwent removal of the variable angle -locking compression plate (lcp) curved condylar plate.It was discovered as broken due to unknown reasons.The surgeon decided to proceed with a total knee revision.The plate was broken through the non-locking portion of the fifth (5th) hole in the shaft closest to the distal end.The patient had an implant of plate and unknown screws on an unknown date.The screws and plate fragments were removed successfully and easily without additional intervention.Procedure was successfully completed with no surgical delay.Procedure had no adverse effect to the patient.Concomitant device: unknown screws (part#unknown, lot#unknown, quantity, unknown).This report is for one (1) 4.5mm va-lcp curved condylar plate/10 hole/230mm/right.This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM VA-LCP CURVED CONDYLAR PLATE/10 HOLE/230MM/RIGHT
Type of Device
IMPLANT,FIXATION DEVICE, CONDYLAR PLATE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
WERK MEZZOVICO (CH)
via cavazz 5
mezzovico 6805
SZ   6805
Manufacturer Contact
kara ditty-bovard
1302 wrights lane east
west chester, PA 19380
6103142063
MDR Report Key8239535
MDR Text Key132788708
Report Number2939274-2019-55791
Device Sequence Number1
Product Code JDP
UDI-Device Identifier10886982042811
UDI-Public(01)10886982042811
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110354
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number02.124.410
Device Catalogue Number02.124.410
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/17/2018
Initial Date FDA Received01/11/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Weight57
-
-