MAUDE Adverse Event Report: CLARET MEDICAL, INC. SENTINEL CEREBRAL PROTECTION SYSTEM; EMBOLIC PROTECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CLARET MEDICAL, INC. SENTINEL CEREBRAL PROTECTION SYSTEM; EMBOLIC PROTECTION DEVICE Back to Search Results
Model Number CMS15-10C
Device Problem Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/14/2018
Event Type  malfunction  
Event Description
It was reported that the distal filter of the sentinel device was punctured.A sentinel embolic protection device had been successfully used in a transcatheter aortic valve replacement (tavr) procedure.After retrieval of the system, the contents of the filter baskets were checked and a puncture in the distal filter was detected.The patients anatomy was reported as not tortuous.Trace paravalvular leak (pvl) was noticed post tavr procedure, and balloon aortic valvuloplasty (bav) was performed to treat this.No patient harm occurred and the patient was reported to be in fine condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENTINEL CEREBRAL PROTECTION SYSTEM
Type of Device
EMBOLIC PROTECTION DEVICE
Manufacturer (Section D)
CLARET MEDICAL, INC.
1745 copperhill parkway
suite 1
santa rosa CA 95403
Manufacturer (Section G)
CLARET MEDICAL, INC.
1745 copperhill parkway
suite 1
santa rosa CA 95403
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key8240093
MDR Text Key132811863
Report Number2134265-2018-64591
Device Sequence Number1
Product Code PUM
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
DEN160043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 01/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2020
Device Model NumberCMS15-10C
Device Catalogue NumberCMS15-10C
Device Lot Number18J12H08
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/17/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/14/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-