Model Number N/A |
Device Problem
Unsealed Device Packaging (1444)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
|
Event Date 08/24/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source, foreign - event occurred in (b)(6).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Device not returned.
|
|
Event Description
|
It was reported that the outer packaging of bone cement was complete, but there was leakage of powder and the inner sterile packaging was opened.Issue noticed before surgery.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.The device was not returned to the manufacturer.Therefore it could not be analyzed.The device manufacturing quality record indicates that the released product met all requirements to perform as intended.According to available data, the most probable root cause is due to packaging manufacturing.Corrective action has been initiated to address reported issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It has been reported that the outer packaging of bone cement was complete, but there was leakage of powder and the inner sterile packaging was opened.Issue noticed before surgery.
|
|