(b)(4).Fse followed up with the customer via telephone to address the reported event.After confirming the reported error, fse instructed the customer to power the instrument off completely, unplug, re-plug, and then perform the buffer change under maintenance.The error subsequently resolved.No further issues were noted.No further action was required by field service.A 13 month complaint history review and service history review for similar complaints was performed for the serial number (b)(4) from 23nov2017 through aware date (b)(4) 2018.There were no similar complaints identified during the search period.The g8 operator's manual under chapter 6- troubleshooting, states the following: 6.3 error messages: when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.140 buffer empty error: the remaining reagent is low.A message will be displayed only if the alarm is set.The operator is instructed to replace the eluent.The most probable cause of the reported event was due to a fault/ failure of the software.
|
The customer reported receiving the "140 buffer empty" error during start up with their g8 analyzer.The customer stated that the hemolysis wash had been changed because it was empty.The customer replaced the buffer per procedure, primed, and rebooted; however, the error reoccurred.Next, technical support (ts) instructed the customer to perform a drain flush, then try to register the wash again.The customer reported receiving the "140 buffer empty" error again, with the gage still at the empty position.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hemoglobin a1c (hba1c).There was no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|