Additional manufacturer narrative: a field service engineer (fse) followed-up with the customer over-the-phone and confirmed that total area on qc was in range after the customer replaced the sample needle assembly.No further action was required by the fse.The g8 analyzer was performing within expected specifications.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) from 01-may-2017 through aware date (b)(4) 2018.There were no other similar complaints identified during the searched period.The g8 variant analysis mode operator's manual under chapter 1, introduction and applications, states the following: 1.8 limitations of the procedure.Total area: dilution studies demonstrate that the assay is linear from a total area of 500 to 4000.However, the optimum total area is 700 to 3000.Chapter 5, maintenance procedures, under section 5.10 provides step-by-step instructions on the sampling needle assembly replacement.Chapter 6, troubleshooting, states the following: 6.3 error messages: when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.200 area low error: three successive results below the lower limit of the total area (50) occur.If the error message is present when sufficient volume of sample is set in the rack, the problem may be caused by an empty reagent (hemolysis & wash solution).Check the remaining volume of hemolysis & wash solution and start the assay again.706 syringe-l error, explanation: operation error in syringe-l.Countermeasure: inspect x1-axis.Inspect syringe-l.Execute smp.Reset.The most probable cause of the reported event was due to failure of the sample needle assembly.(b)(4).
|