• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GIVEN IMAGING LTD., YOQNEAM BRAVO; ELECTRODE, PH, STOMACH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GIVEN IMAGING LTD., YOQNEAM BRAVO; ELECTRODE, PH, STOMACH Back to Search Results
Model Number FGS-0313
Device Problem Loss of or Failure to Bond (1068)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, they had a capsule which failed to attach.The capsule will be returned for investigation.There was no reported patient outcome.
 
Manufacturer Narrative
Additional information: (first name, last name).If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, they had a capsule which failed to attach.There was no harm to the patient, no intervention was required, and no repeat procedure was performed as they used another capsule to complete the procedure.There was nothing unusual about the patient or the procedure and an endoscopy had been performed prior to the procedure and showed the esophagus to be normal.No lubrication was used to facilitate placement of the capsule and the delivery system and capsule will be returned for investigation.
 
Manufacturer Narrative
Additional information: evaluation summary: this report is based on information provided by medtronic investigation personnel and the sample that arrived.One bravo capsule and one bravo delivery device were received for evaluation.A review of the product expiration date discovered this product was used before the expiration date.The returned sample met specification as received by medtronic.The visual inspection found no notable conditions.The customer reported bravo fail to attach to patient's esophagus.The investigation of the returned equipment did not identify anything that would have caused or contributed to the reported event.The investigation found the device to function normally and within specifications.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRAVO
Type of Device
ELECTRODE, PH, STOMACH
Manufacturer (Section D)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
Manufacturer (Section G)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
Manufacturer Contact
amy beeman
15 hampshire street
mansfield,ma, MN 02048
7632104064
MDR Report Key8244430
MDR Text Key132988051
Report Number9710107-2019-00021
Device Sequence Number1
Product Code FFT
UDI-Device Identifier07290101361688
UDI-Public07290101361688
Combination Product (y/n)N
PMA/PMN Number
K102543
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 04/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/20/2018
Device Model NumberFGS-0313
Device Catalogue NumberFGS-0313
Device Lot Number38209Q
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2019
Initial Date Manufacturer Received 12/17/2018
Initial Date FDA Received01/14/2019
Supplement Dates Manufacturer Received12/27/2018
01/31/2019
04/07/2019
Supplement Dates FDA Received01/22/2019
02/27/2019
04/24/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-