Model Number FEM12060
Device Problems Positioning Failure (1158); Material Perforation (2205); Misfire (2532)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/03/2018
Event Type  malfunction  
Manufacturer Narrative
Manufacturing review: based on the information available and the investigation performed, there is no indication that the product failed to meet its specification prior to shipment.The lot records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.The review shows that no remarkable incidents occurred during the manufacturing process.No relevant manufacturing process changes were implemented, that could have led to the event reported.Investigation summary: a stent graft delivery system was returned.Based on the evaluation of the returned sample the reported deployment issue could be confirmed.The stent graft was found partially released and the outer sheath was found to be elongated which indicated that increased friction affected the delivery system during attempt of stent graft deployment.Furthermore, the outer sheath was found perforated by stent graft strut, which made a complete stent graft deployment impossible.No indication was found for manufacturing related issues.Based on the information available and the evaluation of the sample, a definite root cause for the reported event could not be determined.Labeling review: in reviewing the labeling supplied with this product it was found that the instructions for use (ifu) sufficiently address the potential risk.The ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device." regarding the anatomy of the placement site the ifu states: "the safety and effectiveness of the device when placed across a tight bend including the terminal cephalic arch or across the elbow joint has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure." regarding the use of accessories the ifu states: "materials required for the fluency® plus endovascular stent graft procedure: (.) introducer sheath with appropriate inner diameter (.)".Furthermore, the ifu states: "prior to loading the endovascular system over a guide wire, both ports must be flushed with sterile saline (¿).Flushing these lumens will also facilitate stent graft deployment.".
 
Event Description
It was reported that during a stent graft deployment procedure in the upper arm via access in the lower arm, the stent graft allegedly failed to deploy.Therefore another device was used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Type of Device
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstrasse 6
karlsruhe 76227
GM   76227
Manufacturer Contact
judith ludwig
1625 w 3rd st.
tempe, AZ 85281
4803032689
MDR Report Key8244694
MDR Text Key133058887
Report Number2020394-2018-02315
Device Sequence Number1
Product Code PFV
UDI-Device Identifier04049519008646
UDI-Public(01)04049519008646
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130029
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/14/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/25/2021
Device Model NumberFEM12060
Device Catalogue NumberFEM12060
Device Lot NumberANCS3323
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/13/2018
Is the Device Single Use? Yes