Catalog Number 14-500100
Device Problem Material Twisted/Bent (2981)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/17/2018
Event Type  malfunction  
Manufacturer Narrative
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
 
Event Description
It was reported that the tip of a pedicle probe was found bent after surgery.There were no reported surgical impacts.
 
Manufacturer Narrative
Udi number: na.Additional information: method, results, and conclusions - the returned device was examined.The probe was bent and the complaint was confirmed.There isn't a known event associated with the damage, so the root cause cannot be determined.Failures of this type are normally attributed to patient factors, off-axis use, or handling and washing errors.A review of the manufacturing records did not identify any issues which would have contributed to this event.
 
Event Description
It was reported that the tip of a pedicle probe was found bent after surgery.There were no reported surgical impacts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THORACIC PEDICLE PROBE
Type of Device
PROBE
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
ashley mcpherson
10225 westmoor dr.
na
westminster, CO 80021
3034437500
MDR Report Key8244810
MDR Text Key132960277
Report Number3012447612-2019-00020
Device Sequence Number1
Product Code HXB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/14/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number14-500100
Device Lot Number601181
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2019
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/01/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/16/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-