• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. FEMORAL TRIAL SLAPHAMMER; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. FEMORAL TRIAL SLAPHAMMER; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Catalog Number 111140
Device Problems Break (1069); Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/17/2018
Event Type  malfunction  
Manufacturer Narrative
If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Not available.
 
Event Description
Multiple items damaged or did not work during one case.Before the surgery during burr status check the handpiece portion failed and the footpedal portion passed.After that swapped out both the end effector and the anspach.The following ee and anspach passed both burr status checks.During the case after burring the femur and half of the tibia the burr stopped and showed a overheat error.So i swapped out anspach sn (b)(4) for a third loaner set that was at the hospital and the burring was completed without any further issues.During the trialing portion of the case the slap hammer femoral extractor broke.Minimal delay to surgery under 5 mins.This pi is for the joint replacement device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL TRIAL SLAPHAMMER
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
juana hiciano
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key8246267
MDR Text Key133080604
Report Number3005985723-2019-00056
Device Sequence Number1
Product Code NPJ
UDI-Device Identifier00848486002732
UDI-Public00848486002732
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/14/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number111140
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received12/17/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-