Model Number 0.850.0009
Device Problem Battery Problem (2885)
Patient Problem No Patient Involvement (2645)
Event Date 10/26/2018
Event Type  malfunction  
Manufacturer Narrative
The unit was returned and an evaluation was completed.The unit was not received for evaluation until 12/26/2018.Upon initial investigation of the returned device on 01/02/2019 it was identified that there was evidence of burning in one of the handsets.This was not initially reported by the user facility.Upon visual inspection it appears that a thermal event occurred.It is not possible to determine the exact sequence of events that led to the thermal event.However, there is evidence that a short occured in the lower battery pack between the cells and printed circuit board.There was a blackening on the top of the lower battery pack and its printed circuit board (pcb).Cells 2 and 3 appear to have been involved in the thermal event, likely by supplying energy into a fault in the board.Additionally there was a slight melting of the handset plastic enclosure.This concludes the investigation.
 
Event Description
It was reported that the office just wants to send the unit in for an evaluation.There was no report of user or patient injury or impact to patient care.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOMAD PRO
Type of Device
EXTRAORAL SOURCE X-RAY SYSTEM
Manufacturer (Section D)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer (Section G)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer Contact
david waters
11727 fruehauf drive
charlotte, NC 28273
MDR Report Key8252508
MDR Text Key133215998
Report Number1017522-2019-00002
Device Sequence Number1
Product Code EHD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081664
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0.850.0009
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/26/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/26/2018
Initial Date FDA Received01/16/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/11/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-