• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - TIJUANA BAJA CARTRIDGE; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - TIJUANA BAJA CARTRIDGE; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE Back to Search Results
Catalog Number 103401
Device Problems Fluid/Blood Leak (1250); Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/20/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).The actual device was not available; however, two photographs of the sample and eight retained samples were evaluated.Visual inspection of the retention samples did not identify any abnormalities that could have contributed to the reported condition.The samples were leak tested under water and a verification (peeling test) on the union of the arterial line was conducted.No abnormalities were identified from the retention samples that could have contributed to the reported condition.The visual inspection of the provided pictures showed the arterial line was detached from the access site red.From the picture analysis, it was not possible to identify any presence of solvent.A device history review revealed no issues that could have caused or contributed to the reported issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the condition was not determined.This issue is being further investigated.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that during patient treatment with a gambro cartridge set an external blood leak occurred due to a detached arterial line.The amount of blood leak was not reported.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARTRIDGE
Type of Device
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
BAXTER HEALTHCARE - TIJUANA BAJA
tijuana baja california
Manufacturer (Section G)
BAXTER HEALTHCARE - TIJUANA BAJA
blvd pacifico 10014
parque industrial pacifico
tijuana baja california CP 22 643
MX   CP 22643
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8252946
MDR Text Key133233721
Report Number8030638-2018-00026
Device Sequence Number1
Product Code FJK
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K070414
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial
Report Date 01/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number103401
Device Lot Number1000202629
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/20/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-