Siemens has completed an investigation of the reported event.The root cause was determined to be defective hardware.The investigation was performed considering complaint description, cs reports, system log files and system history.The log file analysis shows multiple messages for table transversal movement during roadmap, but no movements were initiated by the user via the control consoles.Because those movements were smaller than 1cm and no actual movements of the table were noticed by the user, this indicates an issue with the table position sensor (potentiometer).In general, during dsa roadmap moving the table longitudinal, transversal, and vertically will cause combined application to reduce exposure (care) position (cross hair) to show in both planes but does not reset roadmap.Table tilt will reset roadmap in b-plane; table rotation will reset a-plane (collimator tracking of table).Any system movement in a single plane will show care position lines in that plane and reset roadmap in that plane opposite plane is unaffected.The potentiometer was replaced by the local service organization.After hardware replacement, no further issues have been reported.The manufacturer is not considering further actions resulting from this event as no systematic error has been recognized.
|