TOSOH BIOSCIENCE, INC. TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/20/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluation by manufacturer: a distributor field service engineer (fse) was dispatched to resolve reported event.Fse performed periodic maintenance, replacing all pm kit parts.Fse successfully completed quality control run within acceptable range.No further action required by field service.The g8 instrument is functioning as expected.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4).There were three (3) similar complaints identified during the searched period, which includes this event.The g8 operator's manual under chapter 1, introduction and applications, states the following: the time from injection of the sample to the time the specific peak elutes off the column is called retention time.The tosoh automated glycohemoglobin analyzer hlc-723g8 software has been written so that each of the expected fractions has a window of acceptable retention times.If the designated peak falls within the expected window, the chromatogram peaks will be properly identified.When a peak elutes at a retention time not within a specified window, an unknown peak (p00) results.If more than one peak elutes at times not specified by the software windows, each is given a sequential p0x title.In order to keep the peaks within their appropriate windows, it may be necessary to change how fast or slow the buffers are moving through the system by changing the pump flow rate.The most probable cause of the reported event was due to instrument needed periodic maintenance.
 
Event Description
A customer reported getting retention time problems with the g8 instrument.A distributor field service engineer (fse) was dispatched to address the reported event, which resulted in delay of results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Manufacturer Narrative
Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.The initial report submitted on 16-jan-2019 to the agency had the incorrect date of 20-sep-2018 under date received by manufacturer.The correct date should have been 17-dec-2018.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH BIOSCIENCE, INC.
3600 gantz road
grove city OH 43123
MDR Report Key8255279
MDR Text Key134590699
Report Number3005529799-2019-00007
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
PMA/PMN Number
K131580
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 01/16/2019,05/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/16/2019
Device Age6 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer01/16/2019
Date Manufacturer Received05/06/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-