(b)(4).Device evaluation by manufacturer: a distributor field service engineer (fse) was dispatched to resolve reported event.Fse performed periodic maintenance, replacing all pm kit parts.Fse successfully completed quality control run within acceptable range.No further action required by field service.The g8 instrument is functioning as expected.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4).There were three (3) similar complaints identified during the searched period, which includes this event.The g8 operator's manual under chapter 1, introduction and applications, states the following: the time from injection of the sample to the time the specific peak elutes off the column is called retention time.The tosoh automated glycohemoglobin analyzer hlc-723g8 software has been written so that each of the expected fractions has a window of acceptable retention times.If the designated peak falls within the expected window, the chromatogram peaks will be properly identified.When a peak elutes at a retention time not within a specified window, an unknown peak (p00) results.If more than one peak elutes at times not specified by the software windows, each is given a sequential p0x title.In order to keep the peaks within their appropriate windows, it may be necessary to change how fast or slow the buffers are moving through the system by changing the pump flow rate.The most probable cause of the reported event was due to instrument needed periodic maintenance.
|