The distributor field service engineer (fse) arrived at the site to address the reported event.The fse decontaminated the g8 analyzer.The fse then replaced the reagents, column, and adjusted the flow rate.No further action was required.The g8 analyzer was performing within specifications.A 13 month complaint history review and service history review for similar complaints was performed for the serial number (b)(4) from 17nov2017 through 17dec2018.There were no similar complaints identified during the search period.The hemoglobin st aia-pack package insert states the following: quality control: in order to monitor and evaluate the precision of the analytical performance, it is recommended that aia-pack hba1c control set be assayed daily.The minimum recommendations for the frequency of running control material are: after calibration, controls are run in order to confirm the calibration curve.Controls should be run if certain service procedures are performed (e.G.Temperature adjustment, sampling mechanism changes, maintenance of the wash probe or detector lamp adjustment or change).After daily maintenance, controls should be run in order to verify the overall performance of the tosoh aia system analyzer.If one or more control sample value(s) is out of the acceptable range, it will be necessary to investigate the validity of the calibration curve before reporting patient results.Standard laboratory procedures should be followed in accordance with the regulatory agency under which the laboratory operates.The g8 operator's manual chapter under chapter 8 - troubleshooting, states the following: adjusting the flow rate - how and why on the tosoh automated glycohemoglobin analyzer hlc-723g8; variant analysis mode it may be necessary to adjust the flow rate because either unidentifiable peaks appear on all the chromatograms or the average retention time for various peaks has changed significantly.The flow rate is changed by changing the flow factor in the instrument.The flow factor is generally 1.00 ml/min.The flow factor should only be adjusted +/- 0.05 of the default factory setting.The most probable cause of the reported event was due to flow rate adjustment was required.
|