The distributor field service engineer (fse) arrived at the site to address the reported event.The fse cleaned the sensors and cleaned and lubricated the mechanism.No further action was required.The g8 analyzer was performing within specifications a 13 month complaint history review and service history review for similar complaints was performed for the serial number (b)(4) from 17nov2017 through 17dec2018.There were no similar complaints identified during the search period.The g8 operator's manual under chapter 6- troubleshooting, states the following: 6.3 error messages: when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.710 z1-axis error: an abnormality occurred in the up and down movement of the sampling needle.If this occurs during a stat assay, check that the container setting (cup or tube) is correctly set.The error also occurs when the sample vial was not recognized as a primary tube, due to the disoriented sample sensor.The most probable cause of the reported event was due to dust obstructing the sensors.
|