• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH BIOSCIENCE, INC. TOSOH HLC-723G8 ANALYZER G8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH BIOSCIENCE, INC. TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/27/2018
Event Type  malfunction  
Manufacturer Narrative
The distributor field service engineer (fse) arrived at the site to address the reported event.The fse cleaned the sensors and cleaned and lubricated the mechanism.No further action was required.The g8 analyzer was performing within specifications a 13 month complaint history review and service history review for similar complaints was performed for the serial number (b)(4) from 17nov2017 through 17dec2018.There were no similar complaints identified during the search period.The g8 operator's manual under chapter 6- troubleshooting, states the following: 6.3 error messages: when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.710 z1-axis error: an abnormality occurred in the up and down movement of the sampling needle.If this occurs during a stat assay, check that the container setting (cup or tube) is correctly set.The error also occurs when the sample vial was not recognized as a primary tube, due to the disoriented sample sensor.The most probable cause of the reported event was due to dust obstructing the sensors.
 
Event Description
It was reported that the customer received "710 z1-axis "errors with their g8 analyzer.The analyzer is down.The distributor field service engineer was dispatched to address the reported event, which resulted in a delay in reporting of patient results for hemoglobin a1c (hba1c).There was no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH BIOSCIENCE, INC.
3600 gatz road
grove city OH 43123
Manufacturer (Section G)
TOSOH BIOSCIENCE, IC.
3600 gantz road
grove city OH 43123
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, 
MDR Report Key8255545
MDR Text Key134925877
Report Number3005529799-2019-00014
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
PMA/PMN Number
K131580
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 01/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/16/2019
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer01/16/2019
Date Manufacturer Received12/17/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/05/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-