Device evaluated by mfr.: returned product consisted of a renegade hi-flo micro-catheter with the fathom guidewire returned stuck in the device.The hub, shaft and tip were microscopically examined.The renegade device showed multiple kinks, bends and stretching located throughout the length of the catheter shaft.It was noticed that the shaft showed a fracture approximately 28.5cm from the hub.The shaft was not completely separated the inner liner was still connected.The guidewire could not be removed due to the damage on the device.The device and the imager were hydrated and the renegade was attempted to be inserted into a 5fr imager angiographic catheter.The renegade would not transcend through the imager due to the damage that was on the renegade device.The inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.No other issues were identified during the product analysis.
|