• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUADRA ASSURA MP ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUADRA ASSURA MP ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3369-40Q
Device Problems Failure to Charge (1085); Delayed Charge Time (2586)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/09/2019
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented to the emergency room after receiving a vibratory alert.Interrogation revealed that during capacitor maintenance, the device exceeded its maximum charge time.Two additional capacitor maintenance tests were performed.Both were within the charge time, but one was significantly longer than the other.No intervention was performed.The patient was stable and will be monitored.
 
Event Description
New information revealed that the device was explanted and replaced.The patient was stable.
 
Manufacturer Narrative
Additional information: d10.Correction: b1 - adverse event, b2 - required intervention to prevent permanent impairment/damage (devices), and h1 - serious injury should have previously been selected in the previous report.The extended charge time was confirmed via reviewing of the device image.Bench testing was performed, and the device subassembly showed normal characteristics.The high voltage capacitors were analyzed and confirmed a failure of a capacitor.Extended charge time was caused by an anomalous hv capacitor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
MDR Report Key8272923
MDR Text Key133933628
Report Number2938836-2019-00548
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508377
UDI-Public05414734508377
Combination Product (y/n)N
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 07/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/23/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2019
Device Model NumberCD3369-40Q
Device Catalogue NumberCD3369-40Q
Device Lot NumberA000042954
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2019
Was the Report Sent to FDA? No
Date Manufacturer Received07/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-