Model Number CD3369-40Q |
Device Problems
Failure to Charge (1085); Delayed Charge Time (2586)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/09/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
It was reported that the patient presented to the emergency room after receiving a vibratory alert.Interrogation revealed that during capacitor maintenance, the device exceeded its maximum charge time.Two additional capacitor maintenance tests were performed.Both were within the charge time, but one was significantly longer than the other.No intervention was performed.The patient was stable and will be monitored.
|
|
Event Description
|
New information revealed that the device was explanted and replaced.The patient was stable.
|
|
Manufacturer Narrative
|
Additional information: d10.Correction: b1 - adverse event, b2 - required intervention to prevent permanent impairment/damage (devices), and h1 - serious injury should have previously been selected in the previous report.The extended charge time was confirmed via reviewing of the device image.Bench testing was performed, and the device subassembly showed normal characteristics.The high voltage capacitors were analyzed and confirmed a failure of a capacitor.Extended charge time was caused by an anomalous hv capacitor.
|
|