Catalog Number 295050-001
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/09/2019
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because this issue did not prevent the freedom driver from performing its life-sustaining functions.In addition, the freedom driver has a redundant power source of onboard batteries.The freedom ac power supply has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The freedom home ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer, a syncardia certified hospital, reported that the freedom ac power supply indicator light is not lighting up and will not charge the onboard batteries.There was no reported adverse patient impact.
 
Manufacturer Narrative
The customer-reported issue of the freedom ac power supply not able to charge freedom onboard batteries was confirmed during the functional evaluation; the ac power supply could not provide enough output to charge onboard batteries installed in a driver.The root cause of the reported issue cannot be determined without physical access to the internal printed circuit board.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.Ce 4632 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM AC POWER SUPPLY
Type of Device
AC POWER SUPPLY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key8277585
MDR Text Key134093403
Report Number3003761017-2019-00011
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 02/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number295050-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/14/2019
Date Manufacturer Received01/09/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age21 YR
-
-